Interventional Radiology - Original Article

Fluid preinjection for microwave ablation in an ex vivo bovine liver model assessed with volumetry in an open MRI system

10.5152/dir.2013.12189

  • Federico Collettini
  • Hendrik Rathke
  • Bernhard Schnackenburg
  • Andreas Thomas
  • Liane Albrecht
  • Britta Suttmeyer
  • Martin Jonczyk
  • Felix Guettler
  • Ulf Teichgräber
  • Thomas Kröncke
  • Bernd Hamm
  • Maximilian de Bucourt

Received Date: 19.12.2012 Accepted Date: 21.03.2013 Diagn Interv Radiol 2013;19(5):427-432

PURPOSE

We aimed to detect possible differences in microwave ablation (MWA) volumes after different fluid preinjections using magnetic resonance imaging (MRI).

MATERIALS AND METHODS

MWA volumes were created in 50 cuboid ex vivo bovine liver specimens (five series: control [no injection], 10 mL water, 10 mL 0.9% NaCl, 10 mL 6% NaCl, and 10 mL 12% NaCl preinjections; n=10 for each series). The operating frequency (915 megahertz), ablation time (7 min), and energy supply (45 watts) were constant. Following MWA, two MR sequences were acquired, and MR volumetry was performed for each sequence.

RESULTS

For both sequences, fluid preinjection did not lead to significant differences in MWA ablation volumes compared to the respective control group (sequence 1: mean MWA volumes ranged from 7.0±1.2 mm [water] to 7.8±1.3 mm [12% NaCl] vs. 7.3±2.1 mm in the control group; sequence 2: mean MWA volumes ranged from 4.9±1.4 mm [12% NaCl] to 5.5±1.9 mm [0.9% NaCl] vs. 4.7±1.6 mm in the control group). The ablation volumes visualized with the two sequences differed significantly in general (P < 0.001) and between the respective groups (control, P ≤ 0.001; water, P < 0.001; 0.9% NaCl, P < 0.001; 6% NaCl, P ≤ 0.001; 12% NaCl, P < 0.001). The volumes determined with sequence 1 were closer to the expected ablation volume of 8 mL compared to those determined with sequence 2.

CONCLUSION

For the fluid qualities and concentrations assessed, there is no evidence that fluid preinjection results in larger coagulation volumes after MWA. Because ablation volumes determined by MRI vary with the sequence used, interventionalists should gain experience in how to interpret postinterventional imaging findings (with the MR scanner, sequences, and parameters used) to accurately estimate the outcome of the interventions they perform.