    Fast kilovoltage-switching dual-energy CT offering lower x-ray dose than single-energy CT for the chest: a quantitative and qualitative comparison study of the two methods of acquisition

    Diagn Interv Radiol 2019;25(3):204-209
    1. Department of Radiology, Yeditepe University School of Medicine, İstanbul, Turkey
    No information available.
    No information available
    Received Date: 11.09.2018
    Accepted Date: 13.11.2018
    PDF
    Cite
    Share
    Request

    ABSTRACT

    PURPOSE

    We aimed to compare the size-specific dose estimates (SSDE), computed tomography (CT) dose indices and image quality parameters of the chest CTs obtained with fast kilovoltage-switching (FKS) dual-energy (DE) CT versus those with single-energy (SE) CT.

    METHODS

    Patients who had chest SECT within the last 6 months were prospectively scanned with chest FKS-DECT. Quantitative comparison was made by calculating the mean SSDE, CTDIvol, contrast, noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) for both acquisitions. Two radiologists evaluated the chest SECT and DECT images qualitatively blinded to the technique used. The paired Student’s t test was utilized for comparing the quantitative and qualitative data. Inter- and intraobserver agreement were also assessed.

    RESULTS

    A total of 42 patients were included. The mean SSDE, CTDIvol, contrast, noise, CNR, and SNR for SECT versus DECT were 12.7±2.2 mGy vs. 9.3±1.2 mGy (P = 0.001), 10.9±2.4 mGy vs. 8±1.2 mGy (P < 0.001), 211.9±44.7 vs. 216.3±59 (P = 0.350), 12.9±2.4 vs. 13.9±3.7 (P = 0.086), 13.5±5.2 vs. 13.3±8.4 (P = 0.548) and 12±3.5 vs. 11.5±3.4 (P = 0.774), respectively. Interobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.89, 0.85, 0.93, and 0.82, respectively; all P < 0.05). Intraobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.80, 0.77, 0.85, and 0.88, respectively; all P < 0.05).

    CONCLUSION

    The mean SSDE of the chest CTs obtained with FKS-DECT were 26.8% lower than those with SECT with significant difference for the objective assessment and there was no significant difference for the subjective assessment of the image qualities, in this series.