ABSTRACT
PURPOSE
A performance comparison of film-screen combination used in mammography was conducted using conventional and new techniques.
RESULTS
Considering the image quality scores from three observers and the radiation doses obtained from the speed measurements, eight combinations were selected as being optimal. Only three of the mammographic film screen combinations in this group were the recommendations of manufacturers.
CONCLUSION
The total performance phantom can be effectively used for the qualitative check of image and provide speed information for mammographic film-screen combinations comparable to sensitometric techniques.