Radiation dose in balloon dacryocystoplasty: a study using Rando® phantoms and thermoluminescent dosimetry
    PDF
    Cite
    Share
    Request
    Interventional Radiology - Original Article
    P: 166-169
    September 2005

    Radiation dose in balloon dacryocystoplasty: a study using Rando® phantoms and thermoluminescent dosimetry

    Diagn Interv Radiol 2005;11(3):166-169
    1. From the Department of Engineering Physics, Ankara University Faculty of Engineering, Ankara, Turkey
    2. Department of Radiology, Gazi University Medical School, Ankara, Turkey
    No information available.
    No information available
    Received Date: 21.02.2005
    Accepted Date: 19.05.2005
    PDF
    Cite
    Share
    Request

    ABSTRACT

    PURPOSE

    The radiation dose to the lens of the eye, skin, thyroid and brain of patients who underwent diagnostic and interventional radiological procedures of the lacrimal drainage system have been measured with thermoluminescent dosimeters (TLD-100 and TLD-700) by using an adult male and female Rando® phantom. All dose values for one second of fluoroscopic exposure and one frame of digital subtraction dacryocystography (DS-DCG) exposure have been obtained individually in the pastero-anterior (PA) and lateral (LAT) projections.

    MATERIALS AND METHODS

    An adult male and female Rando® phantom was used instead of the patients. The procedures were performed by using an Advantx AFM C-arm unit coupled with a DX Hiline digital image acquisition and processing system. The 6-inch mode of a triplefield image intensi.er (II) was used, with a circular collimation of the same or a slightly smaller size. Two different lithium floride (LiF) thermoluminescent chips were used for absorbed dose measurements: TLD700, approximately 4.5 mm in diameter and 0.9 mm in thickness; TLD100, 3.7 mm square and 0.9 mm in thickness.

    RESULTS

    The average values of absorbed doses (lens of the eye, skin, thyroid and brain) measured separately with TLD100 and TLD700 dosimeters after irradiation of the male and female Rando® phantom are presented for LAT and PA projections.

    CONCLUSION

    This study suggests that useful information for dose determination can be obtained by use of the radiation dose to the lens of the eye, thyroid and brain received during radiological procedure of the LDS for one frame of DS-DCG and one second of fluoroscopy.

    Keywords: radiation dose, lacrimal duct obstruction, interventional radiology